Microscope-integrated intraoperative OCT with electrically tunable focus and heads-up display for imaging of
Yuankai K Tao1, Sunil K Srivastava2, Justis P Ehlers2
1Ophthalmic Imaging Center, Cole Eye Institute, Cleveland Clinic, Cleveland, Ohio 44195, USA ; Department of Ophthalmic Research, Cole Eye Institute, Cleveland Clinic, Cleveland, Ohio 44195, USA ; Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio 44195, USA.
Abstract:
We present novel optical and mechanical designs for a microscope-integrated intraoperative optical coherence tomography (iOCT) system with enhanced function and ergonomics for visualization of ophthalmic surgical maneuvers. Integration of an electrically tunable lens allows rapid focal plane adjustment and iOCT imaging of both anterior and posterior segment tissue microstructures while maintaining parfocality with the ophthalmic surgical microscope. We demonstrate novel visualization of instrument positions relative to tissue layers of interest as colormap overlays onto en face OCT data, which may provide integrative display of volumetric information during surgical maneuvers. Finally, we implement a heads-up display system to provide real-time feedback as proof-of-principle for iOCT-guided ophthalmic surgery.