Fully integrated CMOS microsystem for electrochemical measurements on 32 × 32 working electrodes at 90 frames per
Joerg Rothe1, Olivier Frey, Alexander Stettler
1ETH Zurich , Mattenstrasse 26, Basel, 4058, Switzerland.
None:
Microelectrode arrays offer the potential to electrochemically monitor concentrations of molecules at high spatial resolution. However, current systems are limited in the number of sensor sites, signal resolution, and throughput. Here, we present a fully integrated complementary metal oxide semiconductor (CMOS) system with an array of 32 × 32 working electrodes to perform electrochemical measurements like amperometry and voltammetry. The array consists of platinum electrodes with a center-to-center distance of 100 microm and electrode diameters of 5 to 50 microm. Currents in the range from 10 microA down to pA can be measured. The current is digitized by sigma-delta converters at a maximum resolution of 13.3 bits. The integrated noise is 220 fA for a bandwidth of 100 Hz, allowing for detection of pA currents. Currents can be continuously acquired at up to 1 kHz bandwidth, or the whole array can be read out rapidly at a frame rate of up to 90 Hz. The results of the electrical characterization meet the requirements of a wide range of electrochemical methods including cyclic voltammograms and amperometric images of high spatial and temporal resolution.
Related Concept Videos
Electrochemical Systems
Interfacial Electrochemical Methods: Overview
Potentiometry: Membrane Electrodes
Microbial Biosensors