Component position alignment with patient-specific jigs in total knee arthroplasty
Terence R Moopanar1, Jeevaka E Amaranath, Rami M Sorial
1Department of Orthopaedics, Nepean Public Hospital, Penrith, New South Wales, Australia.
ANZ Journal of Surgery
|June 20, 2014
Summary
Patient-specific jigs (PSJ) offer accurate component alignment in total knee arthroplasty (TKA), improving rotational and coronal alignment. Further refinement of tibial guides can enhance PSJ reproducibility and accuracy in TKA procedures.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
- Medical Imaging
Background:
- Improper mechanical alignment and rotational axis are common causes of implant failure in total knee arthroplasty (TKA).
- Conventional and computer-assisted TKA methods have limitations.
- Patient-specific jigs (PSJ) present a potential alternative for TKA component alignment.
Purpose of the Study:
- To evaluate the accuracy and reproducibility of component orientation in TKA using patient-specific jigs (PSJ).
- To compare the efficacy of PSJ with conventional and computer-assisted surgery in TKA.
Main Methods:
- A prospective case series of 261 patients undergoing TKA for osteoarthritis was conducted.
- Preoperative MRI and specialized software were used to generate patient-specific femoral and tibial jigs.
- Postoperative computed tomography (CT) was utilized to measure component alignment.
Main Results:
- 96.2% of patients achieved target femoral rotational alignment.
- 92.7% of tibial coronal alignment was within ±3 degrees of the tibial medullary axis.
- 76.6% of posterior tibial slopes and 81.2% of overall mechanical axes were within acceptable ranges.
Conclusions:
- Patient-specific jigs demonstrate good accuracy for component alignment in TKA compared to conventional methods.
- Improvements in tibial alignment guides are needed to further enhance PSJ accuracy and reproducibility.