Clinical implementation of MRI of joint arthroplasty
Matthew F Koff1, Parina Shah, Hollis G Potter
11 All authors: Department of Radiology and Imaging, MRI Laboratory, Hospital for Special Surgery, 535 E 70th St, New York, NY 10021.
Objective:
The purpose of this article is to explain the basic physics of imaging patients with metal implants, explain conflicting information regarding MRI scanning of "MR Conditional" devices, and relate our experience of scanning total joint arthroplasty (TJA) at our institution.
Conclusion:
MRI near TJA is effective with appropriate imaging protocols and standardized safety precautions. Strict adherence to MR Conditional labeling may preclude broad use of MRI for TJA assessment.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...