Detection of small periprosthetic bone defects after total knee arthroplasty
Related Experiment Video
Updated: Apr 27, 2026

Novel Diagnostics in Revision Arthroplasty: Implant Sonication and Multiplex Polymerase Chain Reaction
Published on: December 3, 2017
Yukihide Minoda1, Taku Yoshida1, Kazutaka Sugimoto2
1Department of Orthopaedic Surgery, Osaka City University Graduate School of Medicine, Osaka, Japan.
Abstract:
Large bone defect around total knee prostheses is among the most critical challenges in revision surgery. However, it is difficult to detect bone defects around a prosthesis in early stage. We compared the efficacy of the detection of small bone defects between fluoroscopically guided plain radiography, CT, MRI, and a novel tomographic technique (tomosynthesis) using the six pig knee models. No bone defects were detected with plain radiography and MRI. The sensitivity and specificity of CT were 61.5% and 64.1%, respectively. The sensitivity and specificity of tomosynthesis were 85.4% and 87.2%, respectively. The radiation dose of tomosynthesis was 6% of that of CT. The cost of tomosynthesis was 28% of that of CT. Tomosynthesis was superior in terms of diagnosis, radiation dose, and cost.