Related Experiment Video
Updated: Apr 27, 2026

Measurement of Pulse Propagation Velocity, Distensibility and Strain in an Abdominal Aortic Aneurysm Mouse Model
Published on: February 23, 2020
Cerebral aneurysm pulsation: do iterative reconstruction methods improve measurement accuracy in vivo?
T Illies1, D Säring2, M Kinoshita3
1From the Departments of Diagnostic and Interventional Neuroradiology (T.I., M.B., J.F.) tillies@uke.de.
Background And Purpose:
Electrocardiogram-gated 4D-CTA is a promising technique allowing new insight into aneurysm pathophysiology and possibly improving risk prediction of cerebral aneurysms. Due to the extremely small pulsational excursions (<0.1 mm in diameter), exact segmentation of the aneurysms is of critical importance. In vitro examinations have shown improvement of the accuracy of vessel delineation by iterative reconstruction methods. We hypothesized that this improvement shows a measurable effect on aneurysm pulsations in vivo.
Materials And Methods:
Ten patients with cerebral aneurysms underwent 4D-CTA. Images were reconstructed with filtered back-projection and iterative reconstruction. The following parameters were compared between both groups: image noise, absolute aneurysm volumes, pulsatility, and sharpness of aneurysm edges.
Results:
In iterative reconstruction images, noise was significantly reduced (mean, 9.8 ± 4.0 Hounsfield units versus 8.0 ± 2.5 Hounsfield units; P = .04), but the sharpness of aneurysm edges just missed statistical significance (mean, 3.50 ± 0.49 mm versus 3.42 ± 0.49 mm; P = .06). Absolute volumes (mean, 456.1 ± 775.2 mm(3) versus 461.7 ± 789.9 mm(3); P = .31) and pulsatility (mean, 1.099 ± 0.088 mm(3) versus 1.095 ± 0.082 mm(3); P = .62) did not show a significant difference between iterative reconstruction and filtered back-projection images.
Conclusions:
CT images reconstructed with iterative reconstruction methods show a tendency toward shorter vessel edges but do not affect absolute aneurysm volumes or pulsatility measurements in vivo.