Characterization of Synthetic Polymers via Matrix Assisted Laser Desorption Ionization Time of Flight MALDI-TOF Mass Spectrometry
Published on: June 10, 2018
Exploration of polyamide structure-property relationships by matrix-assisted laser desorption/ionization
Caroline Barrère1, Majed Rejaibi, Aurélien Curat
1Normandie Université, COBRA, UMR 6014 and FR3038, Université de Rouen; INSA de Rouen; CNRS, IRCOF, 1 rue Tesnière, 76821, Mont-Saint-Aignan Cedex, France.
Rationale:
Polyamides (PA) are among the most used classes of polymers because of their attractive properties. Depending on the nature and proportion of the co-monomers used for their synthesis, they can exhibit a very large range of melting temperatures (Tm ). This study aims at the correlation of data from mass spectrometry (MS) with differential scanning calorimetry (DSC) and X-ray diffraction analyses to relate molecular structure to physical properties such as melting temperature, enthalpy change and crystallinity rate.
Methods:
Six different PA copolymers with molecular weights around 3500 g mol(-1) were synthesized with varying proportions of different co-monomers (amino-acid AB/di-amine AA/di-acid BB). Their melting temperature, enthalpy change and crystallinity rate were measured by DSC and X-ray diffraction. Their structural characterization was carried out by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF-MS). Because of the poor solubility of PA, a solvent-free sample preparation strategy was used with 2,5-dihydroxybenzoic acid (2,5-DHB) as the matrix and sodium iodide as the cationizing agent.
Results:
The different proportions of the repeating unit types led to the formation of PA with melting temperatures ranging from 115°C to 185°C. The structural characterization of these samples by MALDI-TOF-MS revealed a collection of different ion distributions with different sequences of repeating units (AA, BB; AB/AA, BB and AB) in different proportions according to the mixture of monomers used in the synthesis. The relative intensities of these ion distributions were related to sample complexity and structure. They were correlated to DSC and X-ray results, to explain the observed physical properties.
Conclusions:
The structural information obtained by MALDI-TOF-MS provided a better understanding of the variation of the PA melting temperature and established a structure-properties relationship. This work will allow future PA designs to be monitored.
More Related Videos
06:16MALDI-ToF MS Method for the Characterization of Synthetic Polymers with Varying Dispersity and End Groups
Published on: October 3, 2025
07:49Matrix-assisted Laser Desorption/Ionization Time of Flight MALDI-TOF Mass Spectrometric Analysis of Intact Proteins Larger than 100 kDa
Published on: September 9, 2013
Related Concept Videos
Matrix-Assisted Laser Desorption Ionization (MALDI)
MALDI-TOF Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrometry of Amines
Tandem Mass Spectrometry
Mass Spectrometry: Amine Fragmentation
In amines, the number of nitrogen atoms affects the mass of the molecular ion, which is described by the nitrogen rule of mass spectrometry. This rule states that a compound containing...