Related Experiment Video
Updated: Apr 27, 2026

In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Comparison of image quality, myocardial perfusion, and left ventricular function between standard imaging and
Andrew J Einstein1, Ron Blankstein2, Howard Andrews3
1Department of Medicine, Cardiology Division, Columbia University Medical Center and New York-Presbyterian Hospital, New York, New York Department of Radiology, Columbia University Medical Center and New York-Presbyterian Hospital, New York, New York andrew.einstein@columbia.edu.
Unlabelled:
SPECT myocardial perfusion imaging plays a central role in coronary artery disease diagnosis, but concerns exist regarding its radiation burden. Compared with standard Anger SPECT (A-SPECT) cameras, new high-efficiency (HE) cameras with specialized collimators and solid-state cadmium-zinc-telluride detectors offer potential to maintain image quality (IQ), while reducing administered activity and thus radiation dose to patients. No previous study has compared IQ, interpretation, total perfusion deficit (TPD), or ejection fraction (EF) in patients receiving both ultra-low-dose (ULD) imaging on an HE SPECT camera and standard low-dose (SLD) A-SPECT imaging.
Methods:
We compared ULD HE SPECT with SLD A-SPECT imaging by dividing the rest dose in 101 patients at 3 sites scheduled to undergo clinical A-SPECT myocardial perfusion imaging using a same day rest-stress (99m)Tc protocol. Patients underwent HE SPECT imaging after an initial approximately 130-MBq (3.5 mCi) dose and SLD-A-SPECT imaging after the remainder of the planned dose. Images were scored visually by 2 masked readers for IQ and summed rest score. TPD and EF were assessed quantitatively.
Results:
Mean activity was 134 MBq (3.62 mCi) for ULD HE SPECT (effective dose, 1.15 mSv) and 278 MBq (7.50 mCi, 2.39 mSv) for SLD A-SPECT. Overall IQ was superior for ULD HE SPECT (P < 0.0001), with twice as many studies graded excellent quality. Extracardiac activity and overall perfusion assessment were similar. Between-method correlations were high for summed rest score (r = 0.87), TPD (r = 0.91), and EF (r = 0.88).
Conclusion:
ULD HE SPECT rest imaging correlates highly with SLD A-SPECT. It has improved image quality, comparable extracardiac activity, and achieves radiation dose reduction to 1 mSv for a single injection.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...