Performance of the Hospira Plum A+ (HB) hyperbaric infusion pump
Background:
We evaluated the Hospira Plum A+ (HB) hyperbaric infusion pump under monoplace and multiplace hyperbaric conditions to test pump flow accuracy.
Methods:
Pump flow accuracy was tested in monoplace and multiplace hyperbaric chambers at different rates, fluid viscosities, pressures and volumes. Output was recorded from the pump (programmed) and from graduated cylinders or syringes (actual). The lead acid battery life was recorded for multiplace trials.
Results:
In monoplace trials to 3.0 atmospheres absolute (atm abs), the pump functioned within the published tolerance of 12.5% at 1 ml/hour (mean deviation +3.3%, range 0.0% to +5.0%) and at faster flow rates (mean deviation -0.4%, range -11.5% to +6.0%). A trial of packed red blood cells (deviation -3.3%) was also within acceptable limits. For all multiplace trials, the pump functioned well within the manufacturer's limits (mean deviation -0.1%, range -0.8% to +0.9%). At the maximum flow rate, the interval to the first battery alarm for two trials was shorter than the duration of a clinical hyperbaric session (first alarm at 82 and 94 minutes). When we examined delivery variances in the compression and decompression phases for monoplace chambers, at 1 ml/hour 25/39 trials (64%) had no measurable infusion volume during compression. Conversely, more than twice the programmed volume was infused during the 10-minute decompression interval (mean 0.40 ml, range 0.32 to 0.60 ml). The tubing compliance effect was also noted, to a lesser degree, in trials at 3 and 5 ml/hour.
Conclusions:
This infusion pump can be useful in hyperbaric medicine departments that treat patients who need intravenous infusions. Tubing compliance may affect fluid volumes delivered by the pump, especially when delivery rates are low. Careful monitoring of patients with low volume infusions during monoplace chamber compression and decompression is advised.
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Hemodialysis II: Procedure and Complications
Hemodialysis III: Nursing Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...