Study of neutron scattering contribution on Hp(10) and H*(10) calibration in the Brazilian National Low Scattering
B M Freitas1, W W Pereira1, K C S Patrão1
1Instituto de Radioproteção e Dosimetria, Av. Salvador Allende, s/n, Recreio, Rio de Janeiro, CEP: 22783-127 RJ, Brazil.
Abstract:
The neutron scattering at the Low Scattering Laboratory of the Brazilian National Neutron Laboratory has been studied using three different methods. The measurements have been done with a traceable standard (241)Am-Be from source-to-detector distances of 0.52-3.00 m. The obtained results with the variation distance methods are in agreement. Measurements with a large shadow cone are not worth for larger distances due to overshadowing. As the quantity required in a calibration is the response of the device being calibrated to the scattered neutron component in order to subtract this from the total response, for these purposes, the distance variation method must be used for each device. To quantify absolutely the scattering contribution on the quantity rates of fluence, Hp(10) and H*(10) in irradiation procedures, a Bonner sphere spectrometer with the shadow cone was employed. The evaluated scattering correction factor value may be employed for a distance of 1.00 m.
More Related Videos
Related Concept Videos
2D NMR: Heteronuclear Single-Quantum Correlation Spectroscopy (HSQC)
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Lab
2D NMR: Overview of Heteronuclear Correlation Techniques