Related Experiment Video
Updated: Apr 27, 2026

Sample Preparation and Transfer Protocol for In-Vacuum Long-Wavelength Crystallography on Beamline I23 at Diamond Light Source
Published on: April 23, 2021
Performance of a beam-multiplexing diamond crystal monochromator at the Linac Coherent Light Source
Diling Zhu1, Yiping Feng1, Stanislav Stoupin2
1Linac Coherent Light Source, SLAC National Accelerator Laboratory, Menlo Park, California 94025, USA.
Abstract:
A double-crystal diamond monochromator was recently implemented at the Linac Coherent Light Source. It enables splitting pulses generated by the free electron laser in the hard x-ray regime and thus allows the simultaneous operations of two instruments. Both monochromator crystals are High-Pressure High-Temperature grown type-IIa diamond crystal plates with the (111) orientation. The first crystal has a thickness of ~100 μm to allow high reflectivity within the Bragg bandwidth and good transmission for the other wavelengths for downstream use. The second crystal is about 300 μm thick and makes the exit beam of the monochromator parallel to the incoming beam with an offset of 600 mm. Here we present details on the monochromator design and its performance.