Note: Optimized circuit for excitation and detection with one pair of electrodes for improved Fourier transform ion
T Chen1, S C Beu2, N K Kaiser3
1Institute of Molecular Biophysics, Florida State University, Tallahassee, Florida 32306, USA.
Abstract:
A conventional Fourier transform-Ion Cyclotron Resonance (ICR) detection cell is azimuthally divided into four equal sections. One pair of opposed electrodes is used for ion cyclotron excitation, and the other pair for ion image charge detection. In this work, we demonstrate that an appropriate electrical circuit facilitates excitation and detection on one pair of opposed electrodes. The new scheme can be used to minimize the number of electrically independent ICR cell electrodes and/or improve the electrode geometry for simultaneously increased ICR signal magnitude and optimal post-excitation radius, which results in higher signal-to-noise ratio and decreased space-charge effects.