Technical Note: exploring the limit for the conversion of energy-subtracted CT number to electron density for
Masatoshi Saito1, Masayoshi Tsukihara2
1Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Niigata University, Niigata 951-8518, Japan.
Purpose:
For accurate tissue inhomogeneity correction in radiotherapy treatment planning, the authors had previously proposed a novel conversion of the energy-subtracted CT number to an electron density (ΔHU-ρe conversion), which provides a single linear relationship between ΔHU and ρe over a wide ρe range. The purpose of this study is to address the limitations of the conversion method with respect to atomic number (Z) by elucidating the role of partial photon interactions in the ΔHU-ρe conversion process.
Methods:
The authors performed numerical analyses of the ΔHU-ρe conversion for 105 human body tissues, as listed in ICRU Report 46, and elementary substances with Z = 1-40. Total and partial attenuation coefficients for these materials were calculated using the XCOM photon cross section database. The effective x-ray energies used to calculate the attenuation were chosen to imitate a dual-source CT scanner operated at 80-140 kV/Sn under well-calibrated and poorly calibrated conditions.
Results:
The accuracy of the resultant calibrated electron density,[Formula: see text], for the ICRU-46 body tissues fully satisfied the IPEM-81 tolerance levels in radiotherapy treatment planning. If a criterion of [Formula: see text]ρe - 1 is assumed to be within ± 2%, the predicted upper limit of Z applicable for the ΔHU-ρe conversion under the well-calibrated condition is Z = 27. In the case of the poorly calibrated condition, the upper limit of Z is approximately 16. The deviation from the ΔHU-ρe linearity for higher Z substances is mainly caused by the anomalous variation in the photoelectric-absorption component.
Conclusions:
Compensation among the three partial components of the photon interactions provides for sufficient linearity of the ΔHU-ρe conversion to be applicable for most human tissues even for poorly conditioned scans in which there exists a large variation of effective x-ray energies owing to beam-hardening effects arising from the mismatch between the sizes of the object and the calibration phantom.
More Related Videos
06:41Author Spotlight: Optimizing Cryo-EM Analysis with CryoSieve for Enhanced Particle Selection Efficiency
Published on: May 10, 2024
07:24Quantitative Atomic-Site Analysis of Functional Dopants/Point Defects in Crystalline Materials by Electron-Channeling-Enhanced Microanalysis
Published on: May 10, 2021
Related Concept Videos
Atomic Nuclei: Nuclear Spin State Population Distribution
Electron Behavior
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the nucleus have less energy,...
Electron Behavior
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
Thomson's e/m Experiment
A particle with charge q, speed v, and mass m enters an area from the top, where the magnetic and electric fields are perpendicular both to the particle's motion and to one another. The magnetic...
Atomic Radii and Effective Nuclear Charge
Nuclear Binding Energy