Water vapor calibration errors in some capnometers: respiratory conventions misunderstood by manufacturers?
1Department of Anesthesia, University of California, San Francisco 94143-0542.
Abstract:
Incorrect calibration has been included in several recently introduced CO2 analyzers. They display a value of "PCO2" internally calculated as FCO2 X Pb rather than FCO2(Pb - 47) where Pb is barometric pressure. This conceptual error appears to have been introduced because new sampling catheter material has become available that effectively removes water vapor before samples reach the sample cell. This seems to have led some manufacturers to assume, incorrectly, that the 47 mmHg factor used to compute PCO2 in patients would no longer be needed. Users can test whether this error is present in an instrument by testing the effect of wet versus dry gases, and make appropriate corrections if the errors are present. Manufacturers should promptly correct this error in all instruments sold previously.
More Related Videos
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Volatilization
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...