Related Experiment Video
Updated: Apr 27, 2026

In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
Hematologic markers better predict left ventricular assist device thrombosis than echocardiographic or pump
Carlo R Bartoli1, Amaninderapal S Ghotra2, Ajay R Pachika2
1Division of Cardiovascular Surgery, Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania, United States.
Background:
Left ventricular assist device (LVAD) thrombosis is a life-threatening complication that remains a major clinical problem. Consensus diagnostic criteria do not exist. We investigated whether hematologic, echocardiographic, or pump parameters reliably change during LVAD thrombosis.
Methods:
A retrospective analysis of 20 consecutive cases of continuous-flow LVAD thrombosis (Thoratec HeartMate II n = 16, HeartWare HVAD n = 4) was performed. Hematologic markers (lactate dehydrogenase, plasma-free hemoglobin, hemoglobin, creatinine), echocardiographic parameters (left ventricular end-systolic and end-diastolic diameter, mitral regurgitation, aortic insufficiency, inflow-cannula velocity), and pump characteristics (speed, power, estimated flow, pulsatility index) were analyzed with one-way repeated measures ANOVA with Tukey post-test or paired Student t-tests.
Results:
Lactate dehydrogenase and plasma-free hemoglobin were significantly (p < 0.05) elevated at admission for LVAD thrombosis. Hemoglobin and creatinine were not significantly different at admission but changed significantly after admission. Left ventricular end-systolic and end-diastolic diameter, mitral regurgitation, aortic insufficiency, inflow-cannula velocity, LVAD speed, power consumption, estimated flow, and pulsatility index were not significantly different at admission for LVAD thrombosis.
Conclusion:
Hematological markers of hemolysis, but not echocardiographic or pump parameters, reliably changed during LVAD thrombosis. Markers of hemolysis are the best early predictors of LVAD thrombosis.