Related Experiment Video
Updated: Apr 27, 2026

An Image Guided Transapical Mitral Valve Leaflet Puncture Model of Controlled Volume Overload from Mitral Regurgitation in the Rat
Published on: May 19, 2020
CoreValve prosthesis causes anterior mitral leaflet perforation resulting in severe mitral regurgitation
Alberto Cozzarin1, Tomás F Cianciulli2, Robert Guidoin3
1Department of Cardiology, Hospital of the Government of the City of Buenos Aires "Dr. Cosme Argerich", Buenos Aires, Argentina.
Abstract:
Percutaneous transcatheter aortic valve replacement (TAVR) has become an alternative to surgical therapy for patients with severe aortic stenosis and high operative risk, but it is associated with specific complications. We report the case of a 72-year-old man who underwent the procedure without complications; however, 45 days after the procedure, he was admitted to the hospital with symptoms of heart failure secondary to severe mitral regurgitation. Necropsy findings showed prosthesis malposition and perforation of the anterior mitral leaflet caused by the contact of the stent of the CoreValve prosthesis (Medtronic, Minneapolis, MN). We discuss TAVR complications, specifically regarding low positioning of the prosthetic valve.
Related Concept Videos
Mitral Valve Prolapse I: Introduction
Mitral Regurgitation I: Introduction
Mitral Stenosis I: Introduction
Mitral Regurgitation III: Medical Management
Mitral Valve Prolapse II: Assessment and Management
Mitral Regurgitation II: Clinical Features and Diagnostic Tests