Related Experiment Video
Updated: Apr 27, 2026

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
[Analysis of the failures of a cemented constrained liner model in patients with a high dislocation risk]
X Gallart1, J C Gomez1, J A Fernández-Valencia1
1Unidad de Cadera, Servicio de Cirugía Ortopédica y Traumatología, ICEMEQ, Hospital Clínic, Universidad de Barcelona, Barcelona, España.
Objective:
To evaluate the short-term results of an ultra high molecular weight polyethylene retentive cup in patients at high risk of dislocation, either primary or revision surgery.
Material And Method:
Retrospective review of 38 cases in order to determine the rate of survival and failure analysis of a constrained cemented cup, with a mean follow-up of 27 months. We studied demographic data, complications, especially re-dislocations of the prosthesis and, also the likely causes of system failure analyzed.
Results:
In 21.05% (8 cases) were primary surgery and 78.95% were revision surgery (30 cases). The overall survival rate by Kaplan-Meier method was 70.7 months. During follow-up 3 patients died due to causes unrelated to surgery and 2 infections occurred. 12 hips had at least two previous surgeries done. It wasn't any case of aseptic loosening. Four patients presented dislocation, all with a 22 mm head (P=.008). Our statistical analysis didn't found relationship between the abduction cup angle and implant failure (P=.22).
Conclusions:
The ultra high molecular weight polyethylene retentive cup evaluated in this series has provided satisfactory short-term results in hip arthroplasty patients at high risk of dislocation.