Related Experiment Video
Updated: Apr 27, 2026

An Open-Source Normothermic Perfusion System Designed for Research Scientists
Published on: July 18, 2025
Quality control within the multicentre perfusion CT study of primary colorectal cancer (PROSPeCT): results of an
Maria Lewis1, Vicky Goh, Shaun Beggs
1Medical Physics Department, Guy's & St. Thomas' NHS Foundation, Trust, London, UK.
Objectives:
To assess the cross-centre consistency of iodine enhancement, contrast-to-noise ratio and radiation dose in a multicentre perfusion CT trial of colorectal cancer.
Materials And Methods:
A cylindrical water phantom containing different iodine inserts was examined on seven CT models in 13 hospitals. The relationship between CT number (Hounsfield units, HU) and iodine concentration (milligrams per millilitre) was established and contrast-to-noise ratios (CNRs) calculated. Radiation doses (CTDIvol, DLP) were compared across all sites.
Results:
There was a linear relationship between CT number and iodine density. Iodine enhancement varied by a factor of at most 1.10, and image noise by at most 1.5 across the study sites. At an iodine concentration of 1 mg ml(-1) and 100 kV, CNRs ranged from 3.6 to 4.8 in the 220-mm phantom and from 1.4 to 1.9 in the 300-mm phantom. Doses varied by a factor of at most 2.4, but remained within study dose constraints. Iterative reconstruction algorithms did not alter iodine enhancement but resulted in reduced image noise by a factor of at most 2.2, allowing a potential dose decrease of at most 80% compared to filtered back projection (FBP).
Conclusions:
Quality control of CT performance across centres indicates that CNR values remain relatively consistent across all sites, giving acceptable image quality within the agreed dose constraints.
Key Points:
Quality control is essential in a multicentre setting to enable CT quantification. CNRs in a body-sized phantom had the recommended value of at least 1.5. CTDIs and DLPs varied by factors of 1.8 and 2.4 respectively.
More Related Videos
09:00Spatial Measurements of Perfusion, Interstitial Fluid Pressure and Liposomes Accumulation in Solid Tumors
Published on: August 18, 2016
10:38Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
Published on: January 16, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...