Related Experiment Video
Updated: Apr 27, 2026

Assessment of Child Anthropometry in a Large Epidemiologic Study
Published on: February 2, 2017
Differences in noninvasive thermometer measurements in the adult emergency department
Ryan P Bodkin1, Nicole M Acquisto2, Joshua M Zwart1
1Department of Emergency Medicine, University of Rochester Medical Center, Rochester, NY.
Purpose:
Detection of accurate temperature in the emergency department (ED) is integral for assessment, treatment, and disposition. The primary objective was to compare temperature measurements from noninvasive temperature devices in the adult ED. The secondary objective was to evaluate the discrepancy between febrile and afebrile patients.
Methods:
This was a prospective observational study of adult patients presenting to the ED. Patients who required a temperature measurement based on standard of care were included. Data collection included oral and temporal artery (TA) temperature measurement taken consecutively. Data were evaluated using the paired Student's t test.
Results:
A total of 100 patients were identified. Mean oral temperature was 37.51°C (SD ±1.25), and mean TA temperature was 37.03°C (SD ±0.94). The mean difference was 0.48°C (SD ±0.8), P < .0001. Overall, 49% of patients had a difference in temperature measurements greater than or equal to 0.5°C. There were 47 febrile patients, determined by a measurement greater than 38°C on oral or TA thermometer. The mean temperature difference in these patients was 0.87°C (SD ±0.85) compared with a mean temperature difference of 0.12°C (SD ±0.55) in the afebrile patients, P < .0001. A total of 57% of fevers recorded by the oral thermometer were not recorded by the TA thermometer.
Conclusions:
There was a statistically significant difference in measured temperatures between oral and TA thermometers and a clinically significant difference in 49% of patients. Febrile patients had a greater discrepancy and variability between noninvasive temperature measurements. Caution should be taken when evaluating temperature measurements with these noninvasive devices.
More Related Videos
04:34Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
08:22Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...