An enhanced chemiluminescent enzyme immunoassay for serum follicle stimulating hormone is described which involves sequential reaction of anti-follicle stimulating hormone antibody immobilised to the inside surface of an opaque microtitre plate with sample, monoclonal anti-alpha thyroid stimulating hormone antibody, and an anti-mouse IgG - horseradish peroxidase conjungate. Bound peroxidase activity was measured using a p-hydroxycinnamic acid enhanced chemiluminescent luminol-hydrogen peroxide reaction. The assay was sensitive (detection limit 0.01 mU/well) precise (intra-assay precision 2.5-8.1%, inter-assay precision 6.7-11.9%) and results obtained with this assay and a competitive radioimmunoassay were in good agreement (correlation coefficient 0.98).