Unexpected overestimation of methotrexate plasma concentrations: analysis of a single center pediatric population
Emilie Guerriero1, Nicolas Simon, Brigitte Nelken
1*Department of Pharmacy, Lille University Hospital; †Department of Antitumoral Pharmacology, Centre Oscar Lambret; ‡EA4481-GRIIOT, Department of Biopharmacy, Galenic and Hospital Pharmacy, Faculty of Pharmacy, Lille Nord-de-France University; §Department of Paediatric Haematology, Jeanne de Flandres Hospital, Lille University Hospital; and ¶Department of Toxicology, Biology and Pathology Center, Lille University Hospital, France.
Background:
At this center, therapeutic drug monitoring of methotrexate (MTX) used to be performed by fluorescence polarization immunoassay (FPIA). We observed an increasing number of unusual high MTX concentrations at 48 and 72 hours during a couple of years. This study aimed to identify the causes of this variation.
Methods:
A retrospective analysis was conducted on 272 patients hospitalized between January 2008 and October 2012. The whole MTX use system was analyzed using Ishikawa's method. The proportion of MTX concentrations ≤0.2 μmole/L at 48 (P48h) and 72 hours (P72h) was recorded and compared between both FPIA and EMITSiemens assays. A χ or a Fisher exact test was used (α = 0.05).
Results:
Because of an announced withdrawal of the FPIA reagent, the method was switched in 2009 to an immunoenzymatic technique (EMITSiemens). Both P48h and P72h dropped significantly after 2009 (P48h: 45% versus 5% and P72h: 91% versus 47%; P < 0.0001). The replacement of the EMITSiemens reagent by the EMITARK Diagnostics reagent in 2012 led to an increase in both P48h and P72h. No significant difference was found in the proportions of MTX ≤0.2 μmole/L concentrations between FPIA and EMITARK Diagnostics at 48 (45% and 40%; P = 0.556) and 72 hours (91% and 100%; P = 0.231). Both internal and external quality control assessments gave regular satisfactory results during the study period. Furthermore, the interassay comparisons that were performed with internal quality controls and spiked serum samples showed similar results at the time of both shifts. The other changes observed in the MTX circuit were not associated with MTX concentration variations.
Conclusions:
The overestimation of the plasma concentration of MTX was concluded to be because of the assay reagent. A further study is consequently necessary to assess the impact of this analytical pitfall on the patients' survival.
More Related Videos
09:04Comprehensive Evaluation of the Effectiveness and Safety of Placenta-Targeted Drug Delivery Using Three Complementary Methods
Published on: September 10, 2018
08:46Dose Uptake of Platinum- and Ruthenium-based Compound Exposure in Zebrafish by Inductively Coupled Plasma Mass Spectrometry with Broader Applications
Published on: April 21, 2022
Related Concept Videos
Pharmacokinetics in Pediatric Patients: Drug Excretion
Pharmacokinetics in Pediatric Patients: Drug Metabolism
Pharmacokinetics in Pediatric Patients: Drug Distribution
Pharmacokinetics in Pediatric Patients: Overview and Drug Absorption
Drug Dosing: Infants and Children
Therapeutic Drug Monitoring: Affecting Factors