Stability of two angular stable locking plates for open wedge high tibial osteotomy (HTO): TomoFix™ versus LOQTEQ®
Wolf Petersen1, Alexander Wall, Thomas Paulin
1Department of Trauma and Orthopaedic Surgery, Martin Luther Hospital, Caspar Theyss Straße 27-31, Grunewald, 14193, Berlin, Germany, w.petersen@mlk-berlin.de.
Introduction:
The aim of this study was to test the mechanical stability of LOQTEQ® HTO plate in comparison to the TomoFix™ internal plate fixator.
Materials And Methods:
An unstable fracture gap model with two standardized polyoxymethylene (POM) cylinders was used for this study. With this model 5 LOQTEQ® HTO plates and 5 TomoFix™ were cyclically loaded. The start load was 100 N and increased by 50 N after each successful completed cycle. Each load step included 250,000 cycles. All tests were performed in accordance with the American Society for Testing and Materials (ASTM) Standard Specification and Test Method for Metallic Bone Plates F382-99(2003)e1.
Results:
The mean fatigue strength of the LOQTEQ® HTO Plate was 155 % of the value gained by the TomoFix™ implant. This difference was statistically significant (t test <0.001). There was also a significant difference in the cycles but both implants survived. The LOQTEQ® HTO Plate system survived 1,520,336.8 (±121,687.7) cycles, whereas the TomoFix™ group passed on average 847,802,600 (±134,660.7) cycles. In both groups, the only failure mode was plate breakage in a shaft hole.
Conclusions:
The results of the present study showed that fatigue strength of the LOQTEQ® HTO plate was significantly higher in comparison to TomoFix™. These results justify the clinical use of the LOQTEQ® HTO plate.
More Related Videos
10:09Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
Related Concept Videos
Wedges
Consider using a wedge to lift a heavy slab. Here, the wedge functions by converting the applied force into a much larger force directed almost perpendicular to the initial force. This...
Stability of structures