Evaluation of in vitro resistance of different 2.0-mm titanium plates on the mandibular angle sectioning
Objective:
The purpose of this study was to compare, by mechanical in vitro testing, a regular 5-hole plate and a long 4-hole plate with a regular 4-hole plate, applied to stabilize a simulated mandibular angle fracture.
Study Design:
The plates from the 2.0-mm titanium-based system were adapted and stabilized passively in the same site in both groups using four screws, 6 mm long. During the resistance-to-load test, the force was applied perpendicular to the occlusal plane at three different points: first molar at the plated side, first molar at the contralateral side, and between the central incisors.
Results:
The regular 5-hole plates and longer 4-hole plates were superior to the regular 4-hole plates. Furthermore, no statistically significant difference was found between the regular 5-hole plates and long 4-hole plate.
Conclusions:
The length and shape of plates did not interfere with the effectiveness to stabilize the fragments.
Objective:
The purpose of this study was to compare, by mechanical in vitro testing, a regular 5-hole plate and a long 4-hole plate with a regular 4-hole plate, applied to stabilize a simulated mandibular angle fracture.Study design The plates from the 2.0-mm titanium-based system were adapted and stabilized passively in the same site in both groups using four screws, 6 mm long. During the resistance-to-load test, the force was applied perpendicular to the occlusal plane at three different points: first molar at the plated side, first molar at the contralateral side, and between the central incisors.
Results:
The regular 5-hole plates and longer 4-hole plates were superior to the regular 4-hole plates. Furthermore, no statistically significant difference was found between the regular 5-hole plates and long 4-hole plate.Conclusions The length and shape of plates did not interfere with the effectiveness to stabilize the fragments.