Agreement of different methods for assessing sleep characteristics: a comparison of two actigraphs, wrist and hip
Melanie Zinkhan1, Klaus Berger2, Sabrina Hense3
1Institute of Clinical Epidemiology, Martin-Luther-University Halle-Wittenberg, Halle, Germany.
Objective:
To assess the agreement of sleep parameters measured by two actigraphs (SOMNOwatch plus, ActiGraph GT3X+) at two different placements (wrist, hip) and of self-reported sleep with polysomnography (PSG).
Methods:
We estimated agreement with PSG for total sleep time (TST), sleep onset latency (SOL), wake after sleep onset (WASO), number of awakenings after sleep onset (NASO), and sleep efficiency (SE%) for 100 participants of the general population, aged 18-75 years by judging mean differences to PSG and intervals of agreement using Bland-Altman plots.
Results:
Mean difference to PSG for TST was 8.3 min (95% confidence intervals [CI] -7.4; 24.1) for SOMNOwatch plus (wrist), 39.8 min (95% CI 24.3; 55.3) for self-report, -79.0 min (95% CI -89.0; -68.9) for SOMNOwatch plus (hip), and -81.1 min (95% CI -91.9; -70.4) for GT3X+ (hip), respectively. The width of intervals of agreement differed with the placement of the devices. Mean differences to PSG were higher for hip-based measurements compared with wrist placement for most parameters.
Conclusions:
Agreement of sleep parameters assessed by actigraphy with PSG differs with the placement of the device and is limited for hip-based measurements. Agreement of self-report with PSG is comparable to that of actigraphy for some parameters.
Related Concept Videos
Sleep-Wake Cycles
NREM Sleep
NREM sleep comprises four progressive stages that seamlessly merge:
Management of Insomnia
Sedatives and Hypnotics Drugs: Miscellaneous Agents
Melatonin congeners like ramelteon (Rozerem) and tasimelteon (Hetlioz) selectively bind to melatonin receptors (MT1 and MT2) and thus mimic the actions of melatonin, a hormone that regulates sleep-wake cycles. Tasimelteon is primarily used for non-24-hour sleep-wake disorder, common in blind patients. They are also used to treat conditions like insomnia...
Sleep Apnea
The condition is more prevalent among...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Sleepwalking and Sleep Talking
Factors that increase the likelihood of sleepwalking include sleep deprivation and alcohol consumption. Contrary to common beliefs, it is safe...