An alternative approach to account for patient organ doses from imaging guidance procedures
Alan P Nelson1, George X Ding2
1Department of Radiation Oncology, Vanderbilt University School of Medicine, Nashville, USA.
Purpose:
To investigate the feasibility of an alternative method of accounting for additional organ doses resulting from image guidance procedures during patient treatment planning through tabulated values based on scan protocol and scan site.
Methods And Materials:
Patient-specific imaging dose to 30 patients resulting from Varian OBI kV-CBCT scans using the Standard Head (17 patients), Low-dose Thorax (8 patients), and Pelvic (5 patients) scan protocols were retrospectively calculated using Monte Carlo methods. Dose dependence on scan location and patient geometry was explored. Patient organ doses were analyzed by using dose-volume histograms and expressed by the mean, minimum dose delivered to 50% of the organ volume, D50. The reported doses are dose-to-medium instead of dose-to-water.
Results:
The organ doses from all patient-specific calculations show predictable and limited ranges across patients. For brain isocenters using Standard Head Scans: Bone: 0.7-1.1 cGy, Brain: 0.2-0.3 cGy, Brainstem: 0.2-0.3 cGy, Skin: 0.3-0.4 cGy, Eye: 0.03-0.3 cGy. For head and neck patients using the Standard Head Scan: Bone: 0.3-0.6 cGy, Parotids: 0.3-0.4 cGy, Spinal Cord: 0.15-0.25 cGy, Thyroid: 0.1-0.25 cGy, Skin: 0.2-0.3 cGy, Trachea-Esophagus: 0.1-0.2 cGy. For chest using Thorax Scans: Bone: 1.1-1.8 cGy, Soft tissue organs (Bowel, Lung, Heart, Kidney, Esophagus, and Spinal Cord): 0.3-0.6 cGy. For abdominal site using Pelvic Scans: Bone: 3.2-4.2 cGy. Soft tissue organs (Bladder, Bowel, Rectum, Prostate, and Skin) D50s fell between 1.2 and 2.2 cGy. Femoral Heads: 2.5-3.4 cGy.
Conclusions:
It is adequate to estimate and account for organ dose by using tabulated values based on scan procedure and site because organ doses from imaging procedures are only modestly dependent upon scan location and body size. Considering the dose variation and magnitude of dose from each scan protocol in comparison to therapeutic doses, this approach provides a simple alternative to account for additional imaging guidance doses during patient treatment planning. Clinicians can use these tabulated values to make informed decisions in selecting the appropriate imaging procedures and imaging frequency during radiotherapy treatment.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...