Related Experiment Video
Updated: Apr 27, 2026

Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
Microcalcifications versus artifacts: initial evaluation of a new ultrasound image processing technique to identify
Priscilla Machado1, John R Eisenbrey1, Barbara Cavanaugh1
1Department of Radiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
Abstract:
A new commercial image processing technique (MicroPure, Toshiba America Medical Systems, Tustin, CA, USA) that identifies breast microcalcifications was evaluated at the time of patients' annual screening mammograms. Twenty women scheduled for annual screening mammography were enrolled in the study. Patients underwent bilateral outer-upper-quadrant real-time dual gray scale ultrasound and MicroPure imaging using an Aplio XG scanner (Toshiba). MicroPure combines non-linear imaging and speckle suppression to mark suspected calcifications as white spots in a blue overlay image. Four independent and blinded readers analyzed digital clips to determine the presence or absence of microcalcifications and artifacts. The presence of microcalcifications determined by readers was not significantly different from that of mammography (p = 0.57). However, the accuracy was low overall (52%) and also in younger women (<50 years, 54%). In conclusion, although microcalcifications can be identified using MicroPure imaging, this method is not currently appropriate for a screening population and should be used in more focused applications.

