Related Experiment Video
Updated: Apr 27, 2026

Multispectral Real-time Fluorescence Imaging for Intraoperative Detection of the Sentinel Lymph Node in Gynecologic Oncology
Published on: October 20, 2010
Preclinical evaluation of robotic-assisted sentinel lymph node fluorescence imaging
Michael A Liss1, Salman Farshchi-Heydari2, Zhengtao Qin3
1Department of Urology, University of California, San Diego, La Jolla, California UCSD Moores Cancer Center. University of California, San Diego, La Jolla, California.
Unlabelled:
An ideal substance to provide convenient and accurate targeting for sentinel lymph node (SLN) mapping during robotic-assisted surgery has yet to be found. We used an animal model to determine the ability of the FireFly camera system to detect fluorescent SLNs after administration of a dual-labeled molecular imaging agent.
Methods:
We injected the footpads of New Zealand White rabbits with 1.7 or 8.4 nmol of tilmanocept labeled with (99m)Tc and a near-infrared fluorophore, IRDye800CW. One and 36 h after injection, popliteal lymph nodes, representing the SLNs, were dissected with the assistance of the FireFly camera system, a fluorescence-capable endoscopic imaging system. After excision of the paraaortic lymph nodes, which represented non-SLNs, we assayed all lymph nodes for radioactivity and fluorescence intensity.
Results:
Fluorescence within all popliteal lymph nodes was easily detected by the FireFly camera system. Fluorescence within the lymph channel could be imaged during the 1-h studies. When compared with the paraaortic lymph nodes, the popliteal lymph nodes retain greater than 95% of the radioactivity at both 1 and 36 h after injection. At both doses (1.7 and 8.4 nmol), the popliteal nodes had higher (P < 0.050) optical fluorescence intensity than the paraaortic nodes at the 1- and 36-h time points.
Conclusion:
The FireFly camera system can easily detect tilmanocept labeled with a near-infrared fluorophore at least 36 h after administration. This ability will permit image acquisition and subsequent verification of fluorescence-labeled SLNs during robotic-assisted surgery.