Mechanical interference between a closure device and the rigid stent of mitral bioprosthesis: a mechanism for closure
Giuseppe Tarantini1, Stefania Rizzo, Marco Mojoli
1aCardiology Clinic bCardiovascular Pathology, Department of Cardiac, Thoracic and Vascular Sciences, University of Padua, Padua, Italy *Giuseppe Tarantini and Stefania Rizzo contributed equally to the writing of this article.
Abstract:
: We report here the mechanical interference between mitral bioprostheses and devices currently adopted for percutaneous paravalvular leak closure as a possible cause for late dislodgement and embolization of successfully implanted and apparently stable closure devices.The images included in this study demonstrate that the topographical relationship of the target leak with the bioprosthetic valve posts may be a major obstacle for device stability after percutaneous paravalvular leak closure procedures.