An intraoperative device to restore femoral offset in total hip arthroplasty
Enxing Xue, Zhen Su, Chengwang Chen
1Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, 109 Xueyuan West Road, Wenzhou 325000, Zhejiang, China. drzhangyu@hotmail.com.
Background:
Leg length discrepancy (LLD) after total hip arthroplasty (THA) can lead to unsatisfactory outcome. Our objective was to design and evaluate a simple and reliable intraoperative device (Length-offset Lever) to minimize leg length discrepancy.
Methods:
This device was used in 51 patients undergoing primary total hip replacements. The leg length discrepancy was measured pre- and postoperatively based on plain radiographs.
Results:
Preoperative radiographic leg length discrepancy averaged 13.5 ± 6.2 mm. Leg length discrepancy showed significant improvement, with a postoperative average of 4.1 ± 2.3 mm (p < 0.0001). There were no complications associated with this device.
Conclusions:
The 'Length-offset Lever' is a useful intraoperative tool to restore anatomic femoral offset and height of femoral head.