Abstract:
A simple, precise, accurate, robust chiral high-performance liquid chromatographic (chiral HPLC) method was developed for estimation of (S)-piperidin-3-amine (S-isomer) in (R)-piperidin-3-amine dihydrochloride (R-AMP). As AMP is a high-melting solid and nonchromophoric compound, development of a suitable chiral method is a challenging task. The proposed chiral HPLC-UV method involves a precolumn derivatization technique with para toluene sulphonyl chloride (PTSC) in the presence of a base to introduce chromophore into analytes. It utilizes chiralpak AD-H column with a simple mobile phase of 0.1% diethyl amine in ethanol with a 0.5 mL/min flow rate. Analytes were monitored by using a UV detector at 228 nm. The resolution between the two enantiomers was more than 4.0. The developed method was validated as per current International Conference on Harmonization (ICH) guidelines.
More Related Videos
Related Concept Videos
Prochirality
Chirality at Nitrogen, Phosphorus, and Sulfur
A consequence of chirality is the need for enantiomeric resolution. While this is theoretically possible for all...
Racemic Mixtures and the Resolution of Enantiomers
High-Performance Liquid Chromatography: Elution Process
¹H NMR Chemical Shift Equivalence: Enantiotopic and Diastereotopic Protons
In chiral compounds such as 2-butanol, replacing the methylene hydrogens at C3 produces a pair of...
Basicity of Heterocyclic Aromatic Amines