Related Experiment Video
Updated: Apr 26, 2026

Fabrication of Electrochemical-DNA Biosensors for the Reagentless Detection of Nucleic Acids, Proteins and Small Molecules
Published on: June 1, 2011
Electrochemical oligonucleotide-based biosensor for the determination of lead ion
Marta Jarczewska1, Ewa Kierzkowska1, Robert Ziółkowski1
1Institute of Biotechnology, Department of Microbioanalytics, Faculty of Chemistry, Warsaw University of Technology, Noakowskiego 3, 00-664 Warsaw, Poland.
Abstract:
The possibility of utilization of gold electrodes modified with short guanine-rich ssDNA probes for determination of Pb(2+) was examined. Interaction between guanine residues and lead ion followed by formation of G-quadruplex structures was confirmed by electrochemical impedance spectroscopy investigations. An external cationic redox label, methylene blue, was employed in voltammetric measurements for analytical signal generation. It was shown that due to the G-quadruplex formation, the oligonucleotides in the recognition layer fold, which enhances the electron transfer between methylene blue and the electrode surface. The MB current signal rises proportionally to the lead ion concentration in the range from 0.05 to 1μmol/L. The developed biosensor demonstrated high selectivity towards Pb(2+) ion, with only minor response towards interfering metal cations. The calculated limit of detection was of 34.7nmol/L. The utilization of the biosensor for Pb(2+) determination in real samples of water was also tested.
Related Concept Videos
Electrodeposition
Electrodeposition can...
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...
Potentiometry: Membrane Electrodes
Extraction: Advanced Methods