Agent consumption with the Zeus® in the automated closed circuit anesthesia mode with O2/air mixtures
Sofie De Cooman, Jan F A Hendrickx1, Philip John Peyton
1Department of Anesthesiology/CCM, OLV Hospital, Aalst, Belgium. jcnwahendrickx@yahoo.com.
Background:
Earlier software versions of the Zeus® (Lübeck, Dräger, Germany) failed to provide true closed circuit anesthesia (CCA) conditions. We examined whether the latest software (SW 4.03 MK 04672-00) achieves this goal.
Methods:
In 8 ASA I-III patients, the CCA mode of the Zeus® was used to maintain the inspired O2 (FIO2) and end-expired sevoflurane % (FAsevo) at 50 and 1.8%, respectively. The fresh gas flow (FGF) of O2 and air and the sevoflurane injection rate (=Vinjsevo, mL liquid sevo/h) were videotaped from the control screen and entered offline into a spreadsheet. Cumulative sevoflurane usage during early wash-in (=0-1 min, CDsevo0-1), late wash-in (=1-5 min, CDsevo1-5), and maintenance (=5-60 min, CDsevo5-60) was calculated, and Vinjsevo between 1 and 60 min was compared with published uptake data.
Results:
FAsevo reached 1.8% within 101 (23) sec. CDsevo0-1 was between 1.24 (0.03) and 3.01(0.25) mL (a range is provided because no absolute Vinjsevo values were displayed once Vinjsevo was > 100 mL/h, which occurred between 15 ± 2 and 46 ± 6 sec). CDsevo1-5 was 0.81 (0.37) mL, and CDsevo5-60 was 4.63 (0.94) mL. The Vinjsevo pattern between 1 and 60 min matched previously published uptake data. Brief high FGF periods were used to maintain the target FIO2, and to refill the reservoir bag after external pressure had been applied to the abdomen; subsequent "spikes" wasted 0.08-0.19 mL and 0.14-0.49 mL sevoflurane (1-3% and 3-9% of total agent usage between 1 and 60 min, respectively).
Conclusion:
Under the conditions specified, the Zeus® approaches CCA conditions so closely that further reductions in agent usage would have minimal economic significance.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Related Concept Videos
Inhalational Anesthetics: Overview
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...