A method for routine analysis of urinary 4,4'-methylenebis (2-chloroaniline) by gas chromatography-electron capture
Akito Takeuchi1, Akira Namera, Norihiro Sakui
1Osaka Occupational Health Service Center, Japan Industrial Safety and Health Association.
Objectives:
The purpose of this study was to develop a simple and cost-effective method for the determination of urinary 4,4'-methylenebis (2-chloroaniline) (MBOCA) by gas chromatography-electron capture detection (GC-ECD) for biological monitoring of occupational exposure to MBOCA.
Methods:
MBOCA was prepared by liquid-liquid extraction after alkaline hydrolysis, derivatized with N-methyl-bis (trifluoroacetamide) and then analyzed using GC-ECD. The proposed method was validated in accordance with the US Food and Drug Administration guidance.
Results:
The calibration curve showed linearity in the range 1-100 µg/l, with a correlation coefficient of >0.999. The limits of detection and quantification were 0.3 µg/l and 1 µg/l, respectively. The recovery was 94-99%. Intraday accuracy, expressed as the deviation from the nominal value, was 90.5-100.3%, and intraday precision, expressed as the relative standard deviation, was 0.3-2.4%. Interday accuracy and precision were 87.8-100.2% and 0.3-4.1%, respectively.
Conclusions:
The proposed method is a simple and cost-effective method suitable for routine analyses and could be useful for biological monitoring of occupational exposure to MBOCA.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Qualitative Analysis
There are two main approaches to qualitative analysis:...