Tissue dissolution by a novel multisonic ultracleaning system and sodium hypochlorite
Markus Haapasalo1, Zhejun Wang1, Ya Shen1
1Division of Endodontics, Department of Oral Biological and Medical Sciences, Faculty of Dentistry, The University of British Columbia, Vancouver, Canada.
The Multisonic Ultracleaning System significantly enhances tissue dissolution compared to conventional endodontic irrigation devices. This novel system offers a faster and more effective method for tissue removal in dental procedures.
Area of Science:
- Endodontics
- Biomaterials Science
- Biomedical Engineering
Background:
- Evaluating novel endodontic irrigation systems is crucial for improving clinical outcomes.
- Conventional irrigation methods face challenges in effectively removing tissue debris.
- The Multisonic Ultracleaning System presents a new approach to endodontic cleaning.
Purpose of the Study:
- To compare the tissue dissolution efficacy of the Multisonic Ultracleaning System against established endodontic irrigation devices.
- To quantify the rate of tissue dissolution achieved by different irrigation systems under varying conditions.
- To assess the impact of sodium hypochlorite concentration and temperature on tissue dissolution.
Main Methods:
- Bovine muscle tissue was subjected to irrigation by four devices: Multisonic Ultracleaning System, Piezon Master 700, EndoVac, and a conventional needle.
- Irrigation was performed using 0.5%, 3%, and 6% sodium hypochlorite (NaOCl) solutions at 21°C and 40°C.
- Tissue mass change and dissolution time were measured to calculate the rate of tissue dissolution (%/s).
Main Results:
- The Multisonic Ultracleaning System demonstrated the highest rate of tissue dissolution across all tested NaOCl concentrations.
- Tissue dissolution rates increased with higher NaOCl concentrations and temperatures for all devices.
- The Multisonic Ultracleaning System's dissolution rate was over 8 times faster than the next most effective device tested.
Conclusions:
- The Multisonic Ultracleaning System significantly outperforms conventional irrigation devices in terms of in vitro tissue dissolution.
- This novel system shows potential for enhanced cleaning efficacy in endodontic treatments.
- Optimizing NaOCl concentration and temperature further improves tissue dissolution rates.
Related Concept Videos
Tissue Homogenization and Cell Lysis
Mechanical methods of tissue homogenization
These methods rely on applying external physical force to disrupt...
Hand hygiene
Hand washing...
Chemical Agents for Microbial Control
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...