Does surface roughness influence the primary stability of acetabular cups? A numerical and experimental biomechanical
Sophie Le Cann1, Alexandre Galland1, Benoît Rosa2
1Aix-Marseille Université, CNRS, ISM UMR 7287, 13288 Marseille Cedex 09, France; APHM, Institut du Mouvement et de l'appareil Locomoteur, Hôpital Sainte Marguerite, 249 et 270, bd Sainte-Marguerite, 13274 Marseille, France.
Abstract:
Most acetabular cups implanted today are press-fit impacted cementless. Anchorage begins with the primary stability given by insertion of a slightly oversized cup. This primary stability is key to obtaining bone ingrowth and secondary stability. We tested the hypothesis that primary stability of the cup is related to surface roughness of the implant, using both an experimental and a numerical models to analyze how three levels of surface roughness (micro, macro and combined) affect the primary stability of the cup. We also investigated the effect of differences in diameter between the cup and its substrate, and of insertion force, on the cups' primary stability. The results of our study show that primary stability depends on the surface roughness of the cup. The presence of macro-roughness on the peripheral ring is found to decrease primary stability; there was excessive abrasion of the substrate, damaging it and leading to poor primary stability. Numerical modeling indicates that oversizing the cup compared to its substrate has an impact on primary stability, as has insertion force.
Related Concept Videos
Shape and Texture of Coarse Aggregate
Abrasion Resistance of Concrete
One such test is the revolving disc test, where three plates...