Related Experiment Video
Updated: Apr 26, 2026

Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
Published on: March 12, 2021
Comparison of 3D turbo spin-echo SPACE sequences with conventional 2D MRI sequences to assess the shoulder joint
Jost Karsten Kloth1, Marianne Winterstein1, Michael Akbar2
1Diagnostic and Interventional Radiology, University Hospital Heidelberg, Im Neuenheimer Feld 110, D-69120 Heidelberg, Germany.
Purpose:
To determine the accuracy and reliability of three-dimensional (3D) T1- and proton density (PD)-weighted turbo spin-echo (TSE) sampling perfection with application-optimized contrasts using different flip-angle evolution (SPACE) compared with conventional 2D sequences in assessment of the shoulder-joint.
Materials And Methods:
Ninety-three subjects were examined on a 3-T MRI system with both conventional 2D-TSE sequences in T1-, T2- and PD-weighting and 3D SPACE sequences in T1- and PD-weighting. All examinations were assessed independently by two reviewers for common pathologies of the shoulder-joint. Agreement between 2D- and 3D-sequences and inter-observer-agreement was evaluated using kappa-statistics.
Results:
Using conventional 2D TSE sequences as standard of reference, sensitivity, specificity, and accuracy values of 3D SPACE were 81.8%, 95.1%, and 93.5% for injuries of the supraspinatus-tendon (SSP), 81.3%, 93.5%, and 91.4% for the cartilage layer and 82.4%, 98.5%, and 97.5% for the long biceps tendon. Concordance between 2D and 3D was almost perfect for tendinopathies of the SSP (κ=0.85), osteoarthritis (κ=1), luxation of the biceps tendon (κ=1) and adjacent bone marrow (κ=0.92). Inter-observer-agreement was generally higher for conventional 2D TSE sequences (κ, 0.23-1.0), when compared to 3D SPACE sequences (κ, -0.33 to 1.0) except for disorders of the long biceps tendon and supraspinatus tendon rupture.
Conclusion:
Because of substantial and almost perfect concordance with conventional 2D TSE sequences for common shoulder pathologies, MRI examination-time can be reduced by nearly 40% (up to 11 min) using 3D-SPACE without loss of information.
Related Concept Videos
Muscles of the Shoulder
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...
Imaging Studies IV: Magnetic Resonance Imaging
Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...