A theoretical comparison of tissue parameter extraction methods for dual energy computed tomography
Jean-Étienne Tremblay1, Stéphane Bedwani1, Hugo Bouchard2
1Département de Radio-Oncologie, Centre hospitalier de l'Université de Montréal (CHUM), 1560 rue Sherbrooke est, Montréal, Québec H2L 4M1, Canada.
Purpose:
To evaluate the reliability of common sinogram-based DECT reconstruction methods for radiotherapy tissue characterization and to evaluate the advantage of combining them with a stoichiometric calibration.
Methods:
The sinogram-based DECT method defined byAlvarez and Macovski ["Energy-selective reconstructions in x-ray computerized tomography," Phys. Med. Biol. 21, 733-744 (1976)] is adapted to the XCOM photon cross sections database and also generalized to a two-material decomposition method. A theoretical framework is developed using a test phantom containing human tissue compositions for comparing the sinogram-based methods and the calibration-based method, being defined as the application of the stoichiometric calibration technique of Bourque et al. ["A stoichiometric calibration method for dual energy computed tomography," Phys. Med. Biol. 59, 2059-2088 (2014)] on monoenergetic images being generated with a sinogram-based method. Applying a bias correction to the sinogram-based method, its performance in extracting human tissue parameters in the presence of noise as well as by altering the photon energy spectrum is compared to the calibration-based method.
Results:
In the absence of noise and without spectrum alteration, the calibration-based method is found to have no benefit on the sinogram-based method. However, the calibration-based method is shown to be potentially more reliable than bias-corrected sinogram-based methods in situations comparable to the clinical environment, where noise is present and the photon energy spectra can differ from what is used during image reconstruction. In determining electron density, the performance of all methods is comparable in the presence of noise only. Moreover, combined with heavy spectrum alteration, the mean errors on electron density are found higher in sinogram-based methods in comparison with the calibration-based method, with 1.2% versus 0.2%. In the presence of significant noise, bias-corrected sinogram-based methods yield mean errors on effective atomic number of about 2.5%, as compared to 0.5% for the calibration-based method. When combined with heavy spectrum alteration, bias-corrected sinogram-based methods can lead to error of up to 4% on the effective atomic number versus 1.8% for the calibration-based method.
Conclusions:
While sinogram-based methods have the advantage of eliminating beam hardening effects, results of this study suggest improvements in the accuracy and reliability of extracting tissue parameters by applying the DECT stoichiometric calibration of Bourqueet al. to monoenergetic images being generated with such DECT reconstruction methods.
More Related Videos
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
07:013D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography