Radiation exposure to hand surgeons' hands: a practical comparison of large and mini C-arm fluoroscopy
Michael M Vosbikian1, Asif M Ilyas1, Derek D Watson1
1Department of Orthopaedic Surgery, Thomas Jefferson University Hospital; the Department of Orthopaedic Surgery, The Rothman Institute at Thomas Jefferson University Hospital; and the Department of Radiology, Nazareth Hospital, Philadelphia, PA.
Purpose:
To determine and compare the radiation exposure to surgeons' hands with large and mini C-arm fluoroscopy in a practical, clinically based model.
Methods:
Two hand surgeons monitored radiation exposure to their hands with a ring dosimeter over a 14-month period using large and mini C-arm fluoroscopic units. One surgeon performed all cases with a large C-arm unit in a hospital setting, and the other performed all cases with mini C-arms in surgical centers. For each case, fluoroscopic time, the output displayed by the unit, radiation by time, and ring dosimeter absorption were recorded and analyzed.
Results:
A total of 160 consecutive cases were reviewed with 71 cases and 89 cases in the large and mini C-arm groups, respectively. The median output displayed by the large C-arm was 0.7 mGy/case, and the median output displayed by the mini C-arm was 10.0 mGy/case. With output as a product of time, the median calculated values were 0.02 mGy/s for the large C-arm group and 0.28 mGy/s for the mini C-arm group. Cumulative ring dosimeter absorption to the surgeons' hands was found to be 380 mrem for 71 cases in the large C-arm group versus 1,000 mrem for 89 cases in the mini C-arm group.
Conclusions:
In our model, the use of the mini C-arm resulted in more than a 10-fold increase in the rate of output and approximately double the dosimeter absorption to the surgeon's hand compared with the large C-arm. Although it has been shown that the mini C-arm produces less radiation scatter, in a practical model, it may not be a safer alternative with respect to the surgeon's hands. Based on these findings, we recommend that surgeons be more aware of radiation exposure risk, know their C-arm unit's specifications, and try to minimize radiation exposure.
Type Of Study/Level Of Evidence:
Prognostic II.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
07:43In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
Published on: July 2, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Radiological Investigation I: X-ray and CT