Evaluation of radiation exposure of medical staff during CT-guided interventions
Nils Rathmann1, Uwe Haeusler2, Patricius Diezler1
1Institute of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, Medical Faculty Mannheim-Heidelberg University, Mannheim, Germany.
Purpose:
The purpose of this prospective study was to investigate absolute radiation exposure values and factors that influence radiation exposure of interventionists during CT-guided interventions (CTGIs). To our knowledge, no data exist regarding the radiation dose to which the interventionist is exposed during these procedures.
Methods:
Absolute radiation dose values from a total of 131 CTGIs were analyzed. Radiation dose values were collected by thermoluminescent dosimeters that were positioned above the lead protection being worn, on the forehead, thyroid, chest, gonads, and right and left hand and foot. The radiation doses were analyzed with respect to the experience level of the person performing the procedure, the degree of difficulty measured on a 4-point Likert scale, the lesion size measured on a 3-point Likert scale, and the CT system used.
Results:
Median whole-body dose was 12 μSv. With the exception of the forehead, all whole-body radiation doses were statistically significantly lower in CTGIs performed using the modern dual-source CT system compared with the 16-slice multi-detector CT. For CTGIs rated as more complex, the radiation exposure of the radiologist performing the procedure was statistically significantly higher, with the exception of the left hand. A statistically significantly lower median whole-body dose was measured for inexperienced compared with experienced radiologists. However, a few dose measurements of more than 1 mSv were found at the right hand.
Conclusions:
Radiation exposure measured during CTGIs is low (<50 μSv). Because the radiation dose was higher in more-complex interventions and for 16-slice multi-detector row CT, inexperienced radiologists should focus on less-complex procedures.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...