Related Experiment Video
Updated: Apr 26, 2026

Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood
Published on: April 16, 2016
Rapid flow through immunoassay for CRP determination based on polyethylene filters modified with ω-aminocellulose
Thomas Elschner1, Friedrich Scholz, Peter Miethe
1Center of Excellence for Polysaccharide Research, Institute of Organic Chemistry and Macromolecular Chemistry, Friedrich Schiller University of Jena, Humboldtstraße 10, D-07743, Jena, Germany.
Abstract:
A novel method for antibody immobilization is discussed and applied in a column based flow through immunoassay system (ABICAP). It is shown that porous polyethylene material can be modified with different ω-aminocellulose carbamates yielding an amino group containing biocompatible support for antibody immobilization. Anti-h CRP antibodies can be bound covalently to the surface via various homobifunctional cross-linkers. The antibody modified filters are validated for the CRP determination in a sandwich ABICAP system. In a 10 min test procedure based on colloidal dye particles, a limit of detection of 5 ng CRP mL(-1) and coefficients of variation of <9.1% are obtained.
More Related Videos
07:53Activated Cross-linked Agarose for the Rapid Development of Affinity Chromatography Resins - Antibody Capture as a Case Study
Published on: August 16, 2019
09:04Covalent Binding of Antibodies to Cellulose Paper Discs and Their Applications in Naked-eye Colorimetric Immunoassays
Published on: October 21, 2016