Related Experiment Video
Updated: Apr 26, 2026

Author Spotlight: Advancing Rapid Detection of Respiratory Pathogens Using Microfluidic Chip
Published on: March 29, 2024
Rapid and amplification-free detection of fish pathogens by utilizing a molecular beacon-based microfluidic system
Yi-Chih Su1, Chih-Hung Wang1, Wen-Hsin Chang1
1Department of Power Mechanical Engineering, National Tsing Hua University, Hsinchu 30013, Taiwan.
Abstract:
Nervous necrosis virus (NNV) and iridovirus are highly infectious pathogens that can cause lethal diseases in various species of fish. These infectious diseases have no effective treatments and the mortality rate is over 80%, which could cause dramatic economic losses in the aquaculture industry. Conventional diagnostic methods of NNV or iridovirus infected fishes, such as virus culture, enzyme-linked immunosorbent assays and nucleic acid assays usually require time-consuming and complex procedures performed by specialized technicians with delicate laboratory facilities. Rapid, simple, accurate and on-site detection of NNV and iridovirus infections would enable timely preventive measures such as immediate sacrifice of infected fishes, and is therefore critically needed for the aquaculture industry. In this study, a microfluidic-based assay that employ magnetic beads conjugated with viral deoxyribonucleic acid (DNA) capturing probes and fluorescent DNA molecular beacons were developed to rapidly detect NNV and iridovirus. Importantly, this new assay was realized in an integrated microfluidic system with a custom-made control system. With this approach, direct and automated NNV and iridovirus detection from infected fishes can be achieved in less than 30 min. Therefore, this molecular-beacon based microfluidic system presents a potentially promising tool for rapid diagnosis of fish pathogens in the field in the future.
More Related Videos
11:09Multiplex Detection of Bacteria in Complex Clinical and Environmental Samples using Oligonucleotide-coupled Fluorescent Microspheres
Published on: October 23, 2011
09:04Foodborne Pathogen Screening Using Magneto-fluorescent Nanosensor: Rapid Detection of E. Coli O157:H7
Published on: September 17, 2017