Failed inferior vena cava filter retrieval by conventional method: Analysis of its causes and retrieval of it by
Eun Cho1, Kyung Jae Lim1, Jeong Hyun Jo1
1Department of Radiology, Dong-A University Hospital, South Korea.
Objectives:
To evaluate the complications of the temporary implanted inferior vena cava (IVC) filter and the feasibility of double-loop technique for removal of complicated IVC filters.
Methods:
From January 2012 to December 2013, a total of 25 patients with IVC filter were referred for IVC filter retrieval. There were 20 Celect®, 3 OptEase®, and 2 Günther-Tulip® filters. All of the patients were evaluated with a pre-procedural CT scan to identify any complications. The IVC filters which had failed to be retrieved by the conventional method were evaluated, and retrieval was attempted with double loop technique.
Results:
Sixteen of 25 (64%) filters had complications; IVC wall penetration (n = 11, 44%), tilted within IVC (n = 6, 24%), embedded struts (n = 3, 12%), and fracture of the strut (n = 1, 4%). The complications were overlapped in five patients. Two of them (8%) had also complained of filter-related pain. The success rate of IVC filter retrieval by double-loop technique was 14/16 (87.5%). There was no major filter retrieval-related complications.
Conclusions:
The double-loop technique is a safe and feasible method for complicated IVC filter retrieval.