Single acquisition electrical property mapping based on relative coil sensitivities: A proof-of-concept demonstration
José P Marques1, Daniel K Sodickson2, Ozlem Ipek3
1Department of Radiology, University of Lausanne, Lausanne, Switzerland.
Purpose:
All methods presented to date to map both conductivity and permittivity rely on multiple acquisitions to compute quantitatively the magnitude of radiofrequency transmit fields, B1+. In this work, we propose a method to compute both conductivity and permittivity based solely on relative receive coil sensitivities ( B1-) that can be obtained in one single measurement without the need to neither explicitly perform transmit/receive phase separation nor make assumptions regarding those phases.
Theory And Methods:
To demonstrate the validity and the noise sensitivity of our method we used electromagnetic finite differences simulations of a 16-channel transceiver array. To experimentally validate our methodology at 7 Tesla, multi compartment phantom data was acquired using a standard 32-channel receive coil system and two-dimensional (2D) and 3D gradient echo acquisition. The reconstructed electric properties were correlated to those measured using dielectric probes.
Results:
The method was demonstrated both in simulations and in phantom data with correlations to both the modeled and bench measurements being close to identity. The noise properties were modeled and understood.
Conclusion:
The proposed methodology allows to quantitatively determine the electrical properties of a sample using any MR contrast, with the only constraint being the need to have 4 or more receive coils and high SNR. Magn Reson Med 74:185-195, 2015. © 2014 Wiley Periodicals, Inc.
More Related Videos
10:22MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
06:17Quantifying the Relative Thickness of Conductive Ferromagnetic Materials Using Detector Coil-Based Pulsed Eddy Current Sensors
Published on: January 16, 2020
Related Concept Videos
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...
Mutual Inductance
When two circuits carrying time-varying currents are close to one another, the magnetic flux through each circuit varies because of the changing current in the other circuit. Consequently, an emf is induced in each circuit by the changing current in the other. Therefore, this type of emf is called...
Electrical Conductivity
In a practical conductor, an applied electric field may be sustained, causing a flow of electrons, which produce a current. The differential form of the current, the current density, is related to the electric field.
More generally, it is related to the force per unit charge, which involves the...
Magnetic Field Due to Two Straight Wires
Galvanometer
The galvanometer consists of two concave-shaped permanent magnets, providing a uniform radial magnetic field in the annular region. In the center, a pivoted coil of fine copper wire is placed in the uniform...
Magnetic Force On Current-Carrying Wires: Example