Related Experiment Video
Updated: Apr 26, 2026

Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Effect of ActiGraph GT3X+ Position and Algorithm Choice on Step Count Accuracy in Older Adults
Sheila M Korpan1, Jenessa L Schafer, Kaylie C Wilson
1School of Physical Therapy, University of Saskatchewan, Saskatoon, Saskatchewan, Canada.
Abstract:
Accelerometer step count accuracy may be affected by gait speed, device positioning, and analysis algorithm selection. We assessed ActiGraph GT3X+ step count accuracy related to device placement and analysis algorithm in older adults with walking aids (n = 13) and without walking aids (n = 22). Participants (81.5 ± 5.0 years of age) completed a timed 100-m walk wearing five GT3X+ monitors (hips, ankles, lumbar spine). Individuals with walking aids had slower gait speeds (0.8 ± 0.20 m/s versus 1.2 ± 0.20 m/s without walking aids, p < .001). Intraclass correlation coefficient values for observed versus monitored steps were highest when ankle placement and the low frequency extension (LFE) algorithm were used (left ankle ICC = .989, right ankle ICC = .998). Using the GT3X+ ankle placement and analyzing data with the LFE algorithm resulted in the most accurate step counts in older adults.

