Intravoxel incoherent motion (IVIM) imaging at different magnetic field strengths: what is feasible?
Anna S Rydhög1, Matthias J P van Osch2, Emelie Lindgren1
1Department of Medical Radiation Physics, Lund University, Lund, Sweden.
Background:
Due to limited SNR the cerebral applications of the intravoxel incoherent motion (IVIM) concept have been sparse. MRI hardware developments have resulted in improved SNR and this may justify a reassessment of IVIM imaging for non-invasive quantification of the cerebral blood volume (CBV) as a first step toward determining the optimal field strength.
Purpose:
To investigate intravoxel incoherent motion imaging for its potential to assess cerebral blood volume (CBV) at three different MRI field strengths.
Materials And Methods:
Four volunteers were scanned twice at 1.5 T, 3 T as well as 7 T. By correcting for field-strength-dependent effects of relaxation, estimates of corrected CBV (cCBV) were obtained in deep gray matter (DGM), frontal gray matter (FGM) and frontal white matter (FWM), using Bayesian analysis. In addition, simulations were performed to facilitate the interpretation of experimental data.
Results:
In DGM, FGM and FWM we obtained cCBV estimates of 2.2 ml/100 ml, 2.7 ml/100 ml, 1.4 ml/100 ml at 1.5 T; 3.7 ml/100 ml, 5.0 ml/100 ml, 3.2 ml/100 ml at 3 T and 15.5 ml/100 ml, 20.3 ml/100 ml, 7.0 ml/100 ml at 7 T.
Conclusion:
Quantitative cCBV values obtained at 1.5 T and 3 T corresponded better to physiological reference values, while 7 T showed the largest deviation from expected values. Simulations of synthetic tissue voxels indicated that the discrepancy at 7 T can partly be explained by SNR issues. Results were generally more repeatable at 7 T (intraclass correlation coefficient, ICC=0.84) than at 1.5 T (ICC=0.68) and 3 T (ICC=0.46).
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Field due to Moving Charges
Consider a point charge moving with a constant velocity. Like the electric field, the magnetic field at any point is directly proportional to the magnitude of the charge and inversely proportional to the square of the distance between the source point and the field point. However, unlike the electric field, the magnetic field is always perpendicular to the plane containing the line...
Imaging Studies VII: Vascular Imaging
Divergence and Curl of Magnetic Field