An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Laser welded versus resistance spot welded bone implants: analysis of the thermal increase and strength
Carlo Fornaini1, Marco Meleti2, Mauro Bonanini2
1Sezione di Odontostomatologia, Azienda Ospedaliero, Universitaria di Parma, Via A. Gramsci 14, 43100 Parma, Italy ; Laser Unit, Dental Department, Faculty of Medicine, University of Liege, Belgium.
Introduction:
The first aim of this "ex vivo split mouth" study was to compare the thermal elevation during the welding process of titanium bars to titanium implants inserted in pig jaws by a thermal camera and two thermocouples. The second aim was to compare the strength of the joints by a traction test with a dynamometer.
Materials And Methods:
Six pigs' jaws were used and three implants were placed on each side of them for a total of 36 fixtures. Twelve bars were connected to the abutments (each bar on three implants) by using, on one side, laser welding and, on the other, resistance spot welding. Temperature variations were recorded by thermocouples and by thermal camera while the strength of the welded joint was analyzed by a traction test.
Results:
For increasing temperature, means were 36.83 and 37.06, standard deviations 1.234 and 1.187, and P value 0.5763 (not significant). For traction test, means were 195.5 and 159.4, standard deviations 2.00 and 2.254, and P value 0.0001 (very significant).
Conclusion:
Laser welding was demonstrated to be able to connect titanium implant abutments without the risk of thermal increase into the bone and with good results in terms of mechanical strength.