Related Experiment Video
Updated: Apr 26, 2026

Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
A new technique for noise reduction at coronary CT angiography with multi-phase data-averaging and non-rigid image
Fuminari Tatsugami1, Toru Higaki, Yuko Nakamura
1Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, 734-8551, Japan, sa104@rg8.so-net.ne.jp.
Objectives:
To investigate the feasibility of a newly developed noise reduction technique at coronary CT angiography (CTA) that uses multi-phase data-averaging and non-rigid image registration.
Methods:
Sixty-five patients underwent coronary CTA with prospective ECG-triggering. The range of the phase window was set at 70-80% of the R-R interval. First, three sets of consecutive volume data at 70%, 75% and 80% of the R-R interval were prepared. Second, we applied non-rigid registration to align the 70% and 80% images to the 75% image. Finally, we performed weighted averaging of the three images and generated a de-noised image. The image noise and contrast-to-noise ratio (CNR) in the proximal coronary arteries between the conventional 75% and the de-noised images were compared. Two radiologists evaluated the image quality using a 5-point scale (1, poor; 5, excellent).
Results:
On de-noised images, mean image noise was significantly lower than on conventional 75% images (18.3 HU ± 2.6 vs. 23.0 HU ± 3.3, P < 0.01) and the CNR was significantly higher (P < 0.01). The mean image quality score for conventional 75% and de-noised images was 3.9 and 4.4, respectively (P < 0.01).
Conclusions:
Our method reduces image noise and improves image quality at coronary CTA.
Key Points:
• We introduce a new method for image noise reduction at cardiac CT. • Multiple data acquisitions of an object and their averaging yield lower noise. • Our method uses multi-phase images reconstructed from unused redundant imaging data. • It reduces image noise by averaging multi-phase images transformed by non-rigid registration. • This method achieves a 20% image noise reduction at cardiac CT.