Morphometric Analyses of Retinal Sections
Published on: February 19, 2012
Reproducibility and repeatability of ganglion cell-inner plexiform layer thickness measurements in healthy subjects
Paolo Carpineto1, Agbeanda Aharrh-Gnama, Vincenzo Ciciarelli
1Ophthalmology Clinic, University 'G. d'Annunzio' Chieti-Pescara, Chieti, Italy.
Purpose:
To assess the reproducibility and repeatability of macular ganglion cell-inner plexiform layer (GC-IPL) thickness measurements in healthy subjects.
Procedures:
In this observational study, 60 healthy eyes were subjected to macular GC-IPL thickness measurements by means of Cirrus™ high-definition optical coherence tomography (Cirrus version 6.0; Carl Zeiss Meditec, Dublin, Calif., USA) by two examiners in two sessions. Average, minimum and 6 sectoral GC-IPL thicknesses were measured. Inter- and intraobserver reproducibility was tested and analyzed by means of the concordance correlation coefficient (CCC). The repeatability of measurements was assessed by the coefficient of repeatability (CR).
Results:
Mean age (±SD) was 29.63 (±5.1) years. The CRs for average GC-IPL thickness were 2.1 and 2.2 µm for the first and the second operator, respectively. Inter- and intraobserver CCCs ranged from 0.91 (95% CI: 0.89-0.93) to 0.98 (95% CI: 0.96-0.99) and from 0.92 (95% CI: 0.88-0.94) to 0.98 (95% CI: 0.97-0.99), respectively.
Conclusions:
GC-IPL thickness measurements in young healthy subjects showed excellent reproducibility and repeatability, especially for average and sectoral GC-IPL thickness measurements.