Diffusion-weighted imaging of the liver: comparison of image quality between monopolar and bipolar acquisition

Andrew B Rosenkrantz1, Christian Geppert, Michael Kiritsy

  • 1Department of Radiology, NYU Langone Medical Center, 550 First Avenue, New York, NY, 10016, USA, Andrew.Rosenkrantz@nyumc.org.

Abdominal Imaging
|August 14, 2014
PubMed
Abstract

Insights

Monopolar diffusion-weighted imaging (DWI) offers superior liver MRI quality at 3T compared to bipolar DWI, demonstrating higher signal-to-noise ratio and better diagnostic clarity. This monopolar technique enhances image quality for liver assessments.

Area of Science:

  • Magnetic Resonance Imaging (MRI)
  • Radiology
  • Medical Imaging Physics

Background:

  • Diffusion-weighted imaging (DWI) is crucial for liver assessment.
  • Optimizing DWI sequence quality at 3T magnetic resonance imaging (MRI) is essential for accurate diagnosis.
  • Monopolar and bipolar DWI sequences have different characteristics affecting image quality.

Purpose of the Study:

  • To compare the image quality of monopolar and bipolar diffusion-weighted imaging (DWI) sequences for liver imaging at 3T.
  • To evaluate signal-to-noise ratio (eSNR), apparent diffusion coefficient (ADC), and subjective image quality between the two DWI techniques.

Main Methods:

  • 32 healthy volunteers and 11 patients underwent liver MRI at 3T.
  • Free-breathing DWI sequences (b-values 0, 400, 800 s/mm²) were acquired using both monopolar and bipolar techniques.
  • Image quality metrics including eSNR, ADC, and radiologist-scored subjective quality were compared.

Main Results:

  • The monopolar sequence showed significantly higher eSNR in both volunteers and patients.
  • Radiologists assigned higher scores for subjective image quality (clarity, conspicuity, diagnostic quality) to the monopolar sequence.
  • No significant difference in respiratory motion artifact was observed between sequences.

Conclusions:

  • Monopolar DWI sequences provide improved liver image quality at 3T compared to bipolar sequences.
  • Higher eSNR is a primary factor contributing to the enhanced image quality with the monopolar technique.
  • The findings suggest monopolar DWI is preferable for liver imaging at 3T.

Related Concept Videos

Imaging Studies IV: Magnetic Resonance Imaging01:27

Imaging Studies IV: Magnetic Resonance Imaging

Introduction:Magnetic Resonance Imaging, or MRI, can include a specialized imaging technique of the urinary system known as Magnetic Resonance Urography (MRU). This radiation-free technique uses strong magnetic fields and radio waves to produce detailed images with the help of a computer. MRU is particularly effective for visualizing fluid-filled structures like the kidneys, ureters, and bladder.Applications of MRI in the Genitourinary SystemKidneys and Ureters: MRI detects tumors, cysts,...
417
Magnetic Resonance Imaging01:24

Magnetic Resonance Imaging

Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
7.5K
Imaging Studies for Cardiovascular System IV: CMRI01:21

Imaging Studies for Cardiovascular System IV: CMRI

Cardiovascular magnetic resonance imaging, or CMRI, is a non-invasive diagnostic test that employs a magnetic field and radiofrequency waves to create precise images of the heart and arteries. It provides comprehensive information about cardiac anatomy, function, perfusion, and tissue characterization without ionizing radiation.IndicationsCMRI diagnoses various heart conditions, including tissue damage from heart attacks, ischemic heart disease, myocarditis, aortic issues (tears, aneurysms,...
529