Low-dose CT for patients with clinically suspected acute appendicitis: optimal strength of sinogram affirmed
Seung Ho Kim1, Jung-Hee Yoon2, Jang Hee Lee2
1Department of Radiology, Inje University College of Medicine, Haeundae Paik Hospital, Busan, Republic of Korea radiresi@gmail.com.
Background:
As there is increased concern over the radiation exposure particularly in adolescents and young adults, computed tomography (CT) dose reduction is needed in the diagnosis of acute appendicitis.
Purpose:
To evaluate the optimal strength of sinogram affirmed iterative reconstruction (SAFIRE) to obtain the best image quality on a 30-mAs applied low-dose CT (LDCT 30mAs) and to compare the diagnostic performances of the LDCT 30mAs with different SAFIRE strengths with that of the 100-mAs applied LDCT (LDCT 100mAs) for the diagnosis of acute appendicitis.
Material And Methods:
A total of 102 consecutive patients (47 men, 55 women; mean age, 41.2 years; range, 15-82 years) with right lower quadrant pain underwent abdominal-pelvic CT, consisting of arterial phase LDCT 100mAs and portal venous phase LDCT30mAs under a fixed 120 kV. LDCT 30mAs images were reconstructed separately with five strength levels (S1-S5). Two blinded radiologists recorded scores for the subjective image quality of the LDCT 30mAs dataset (S0-S5) and confidence scores for the diagnosis of acute appendicitis on each dataset and LDCT 100mAs. CT image noise was measured for each set.
Results:
The study population consisted of 58 patients with confirmed appendicitis and 44 without appendicitis. There was no significant difference in diagnostic performance between LDCT 100mAs and LDCT 30mAs with any strength for both readers (AUC for reader 1, LDCT 30mAs with S0-S5 = 0.97, LDCT 100mAs = 0.93, P = 0.0936; for reader 2, LDCT 30mAs with S0-S5 = 0.96, LDCT 100mAs = 0.97, P = 0.128). The measured noise decreased as the strength increased from S0 to S5 (mean, 20.8 > 17.7 > 15.6 > 13.5 > 11.5 > 9.5, P < 0.0001). However, overall subjective image quality on S3 was better than the other strengths for both readers (S0 < S1 < S2 < S3 > S4 > S5, P < 0.0001).
Conclusion:
Although measured noise declined as SAFIRE strength increased, S3 seems optimal for the best subjective image quality on LDCT 30mAs. The diagnostic performance of LDCT 30mAs with any strength is comparable to that of LDCT 100mAs for the diagnosis of acute appendicitis.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Appendicitis-II: Diagnostic Studies and Management
Diagnosing Appendicitis
It requires a multifaceted approach, starting with a detailed physical examination to pinpoint the location and nature of the pain and identify any associated symptoms. Laboratory tests play a crucial role. A complete Blood Count (CBC) typically reveals leukocytosis (an increased number of...
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT